MedicineBiomedical EngineeringSensorsUltrasonographyUltrasoundDiagnostic UltrasonographyInterferometryUltrasonicsUltrasound ImagingUltrasound InstrumentationUltrasound PhysicsInterferenceMach-zehnder InterferometersResonanceRing Resonator SensorsSignalsUltrasonic SensingUltrasonic SensorsVibrational Modes
on the circle arc periodically traced in the plane of the two orthogonal interrogator voltages, as the principal sensor signal. Interrogation is demonstrated for two sensors with membrane vibrational modes at 1.3 and 0.77 MHz, by applying continuous wave ultrasound in a wide pressure range. Ultrasound is detected at a pressure as low as 1.2 Pa. Two optical path differences (OPDs) of the MZI are used. Thus, different interference conditions of the optical signals are defined, leading to a higher apparent sensitivity for the larger OPD, which is accompanied by a weaker signal, however. Independent measurements using the modulation method yield a resonance modulation per unit of pressure of 21.4 fm/Pa (sensor #1) and 103.8 fm/Pa (sensor #2).