Journal of Cardiac Surgery · 2017 · 12 citations · 7 references
MedicineCardiologySurgeryBiomedical ImagingCardiothoracic SurgeryDiagnostic ImagingCardiothoracic ImagingCardiovascular MonitoringInfrared InstrumentationIntravascular ImagingPostoperative Cardiac MonitoringPostoperative MonitoringInfrared CamerasInfrared ThermographyIntraoperative Cardiac AssessmentPatient Temperature MonitoringThermal ImagingThermal Imaging AssessmentThermal Infrared ImagingThermography
We describe the intraoperative non-invasive use of an infrared (IR) camera to monitor Del Nido cardioplegia delivery in patients undergoing cardiac surgery. Thermal pictures were taken pre- and post-cardioplegia and at timed points after arrest, and compared to readings from a transseptal temperature probe. There was good concordance between the transseptal probe and the IR camera temperature readings. This non-invasive technique, which assesses cardioplegic distribution, may help to determine when additional doses of Del Nido cardioplegia are required during periods of cardioplegic arrest.
7
Inadequate Cardioplegic Protection with Obstructed Coronary Arteries
Colin J. Hilton, William Teubl, Michael Acker et al. · The Annals of Thoracic Surgery · 1979 · 167 citations · Full text
Medicine, Cardiology, Surgery +9
Studies of retrograde cardioplegia
Marshall T. Partington, Christophe Acar, Gerald D. Buckberg et al. · Journal of Thoracic and Cardiovascular Surgery · 1989 · 101 citations