Publication | Open Access
Mechanical fatigue resistance of an implantable branched lead system for a distributed set of longitudinal intrafascicular electrodes
21
Citations
39
References
2017
Year
These results demonstrate that the branched lead of this implantable neural interface system has sufficient mechanical fatigue resistance to withstand strain profiles anticipated when the system is implanted in an arm. The novel test setups and paradigms may be useful in testing other lead systems.