Concepedia

Publication | Closed Access

Development and Validation of Quantitative Ultraperformance Liquid Chromatography–Tandem Mass Spectrometry Assay for Anticoagulant Rodenticides in Liver

28

Citations

20

References

2017

Year

Abstract

Anticoagulant rodenticides (ARs) are used to control rodent populations; however, exposure to nontarget animals occurs. A sensitive and rugged quantitative method was developed, optimized, and validated for eight ARs in liver. Target analytes comprised two chemical classes: hydroxycoumarins (warfarin, coumachlor, dicoumarol, bromadiolone, brodifacoum, and difethialone) and indanediones (diphacinone and chlorophacinone). In this method, liver extracts were cleaned using dispersive solid phase extraction (d-SPE) to remove matrix interferences and analyzed by reverse phase ultraperformance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS). Electrospray ionization in negative ion mode combined with multiple reaction monitoring (MRM) using a triple quadrupole mass spectrometer provided simultaneous confirmation and quantitation. Detection limits spanned 0.75-25 ng/g, and lower quantitation limits were established as 50 ng/g. Interassay method accuracy ranged from 92 to 110% across the analytical range (50-2500 ng/g) using matrix-matched calibrants with good repeatability (relative standard deviations 2-16%). Successful method transfer to another laboratory utilizing an Orbitrap mass analyzer, providing high mass accuracy, was assessed by good method reproducibility during blinded study analyses (6-29%; Horwitz ratios (HORRAT) ≤ 1.5).