Publication | Open Access
Evaluation of RF Heating on Hip Joint Implant in Phantom during MRI Examinations
30
Citations
13
References
2010
Year
RF heating was confirmed to take place at both ends of the implants in spite of their different shapes. The maximum temperature rise was observed at the tip where there is large curvature. The value was found to depend on physical properties of the implant materials. The discrepancy between experimental and simulated temperature rises was presumed to be the result of an incomplete model for the titanium implant.