The International Journal of Artificial Organs · 2017 · 13 citations · 9 references
MedicineSurgeryCardiothoracic SurgeryHemodynamic ManagementRenal Replacement TherapyContinuous Renal Replacement TherapyExtracorporeal CirculationExtracorporeal Life SupportExtracorporeal Membrane OxygenationAdult Extracorporeal Membrane OxygenationBlood FlowExtracorporeal Circuit MonitoringExtracorporeal Membrane Oxygenation MonitoringMembrane Oxygenation
PURPOSE: This study aimed to assess a new approach combining venoarterial (VA) extracorporeal membrane oxygenation (ECMO) and continuous renal replacement therapy (CRRT) in adults, while monitoring CRRT circuit pressures. METHODS: The inlet and outlet of the CRRT circuit were connected to preoxygenator port and postoxygenator port, respectively. Then, complications and CRRT circuit pressures were evaluated. RESULTS: 7 patients underwent combined VA-ECMO and CRRT; 16 filters were used. CRRT blood flow ranged from 150 to 200 mL/min; the CRRT to ECMO blood flow ratio was <0.1. The CRRT pressures at treatment initiation were normal. No complications were reported. CONCLUSIONS: This approach combining VA-ECMO and CRRT in adults did not compromise the accuracy of pressure monitoring systems for CRRT circuit function, and caused no complications. Hence, it may be a feasible method for performing combined VA-ECMO and CRRT in adults.
9
A Multicenter International Survey of Renal Supportive Therapy During ECMO
Geoffrey M. Fleming, David J. Askenazi, Brian C. Bridges et al. · ASAIO Journal · 2012 · 178 citations
Medicine, Nephrology, Hemodialysis +12