Evaluations of the setup discrepancy between BrainLAB 6D ExacTrac and cone-beam computed tomography used with the imaging guidance system Novalis-Tx for intracranial stereotactic radiosurgery
23
Citations
17
References
2017
Year
Computed TomographySurgeryTreatment VerificationNeuro-oncologyCt ScanRadiation Therapy PlanningNeurologyRadiation ImagingRadiologyHealth SciencesGuidance System Novalis-txMedical ImagingNeuroimagingImage GuidanceRadiologic ImagingMri-guided Radiation TherapyExactrac SystemSetup DiscrepancyDiagnostic NeuroradiologyBrainlab 6Brainlab 6DCraniofacial SurgeryMedicine
The objective of this study was to evaluate the setup discrepancy between BrainLAB 6 degree-of-freedom (6D) ExacTrac and cone-beam computed tomography (CBCT) used with the imaging guidance system Novalis Tx for intracranial stereotactic radiosurgery. We included 107 consecutive patients for whom white stereotactic head frame masks (R408; Clarity Medical Products, Newark, OH) were used to fix the head during intracranial stereotactic radiosurgery, between August 2012 and July 2016. The patients were immobilized in the same state for both the verification image using 6D ExacTrac and online 3D CBCT. In addition, after radiation treatment, registration between the computed tomography simulation images and the CBCT images was performed with offline 6D fusion in an offline review. The root-mean-square of the difference in the translational dimensions between the ExacTrac system and CBCT was <1.01 mm for online matching and <1.10 mm for offline matching. Furthermore, the root-mean-square of the difference in the rotational dimensions between the ExacTrac system and the CBCT were <0.82° for online matching and <0.95° for offline matching. It was concluded that while the discrepancies in residual setup errors between the ExacTrac 6D X-ray and the CBCT were minor, they should not be ignored.
| Year | Citations | |
|---|---|---|
Page 1
Page 1