<title>CAMILA: Infrared Camera/Spectrograph for OAN-SPM</title>

Luis Carrasco, Elfego Ruiz Schneider, Lorenzo S. Leija, Michael F. Skrutskie, Michael R. Meyer, Pablo D. Sotelo, Paula Barbosa, Leonel Gutiérrez, Arturo I. Iriarte Valverde, Francisco J. Cobos Duenas,

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 1994 · 27 citations · 0 references

Concepts

Abstract

The development of the IR camera and spectrograph (CAMILA) is described. It is based on a NICMOS 3 HgCdTe detector developed by Rockwell with a spectral response of 1 to 2.5 micrometers . The initial configuration of the system was recently concluded and consists of the following components: detector cryostat, detector control electronics, low noise preamplifiers, detector-PC interface, operating system and optics. The characterization of the electronics and the science grade chip are presented. The complete optical configuration allows the following modes of operation: direct imaging (12 filter positions), polarimetry and spectroscopy on three dispersion modes (low, medium, and high resolution). Preliminary spectroscopic results at the H band with R equals 1500 are presented. The project is a collaborative effort of groups from IAUNAM and UMASS (Amherst) and will be used mainly at the 2.1-m telescope of San Pedro Martir, B.C. (Mexico).