Journal of the Optical Society of America A · 2017 · 28 citations · 11 references
Natural SciencesElectrical EngineeringAnalytical ChemistryPhotonicsSpectroscopyAnalytical CalibrationCalibrationOptical SpectroscopyPrecision SpectroscopySpectral CalibrationSpectral Data ProcessingAccuracyBandwidthCharge-coupled Device CalibrationLine Profile AnalysisLine Profile FittingSpectral Line CalibrationSpectrometer CalibrationWavelength Calibration
Wavelength calibration is an important step in charge-coupled device (CCD) spectrometers. In this paper, an accurate calibration method is proposed. A model of a line profile spectrum is built at the beginning, followed by noise reduction, bandwidth correction, and automatic peak-seeking treatment. Experimental tests are conducted on the USB4000 spectrometer with a mercury-argon calibration light source. Compared with the traditional method, the results show that this wavelength calibration procedure obtains higher accuracy and the deviations are within 0.1 nm.
11
Computer fitting of Voigt profiles to Raman lines
Tom Sundius · Journal of Raman Spectroscopy · 1973 · 82 citations
Natural Sciences, Chemistry, Physics +15