Journal of Physics Conference Series · 2015 · 10 citations · 0 references
Natural SciencesElectrical EngineeringPhysicsAtomic PhysicsMicrowave EngineeringQuantum MechanicsAtomic ClocksAtomic ExcitationFrequency MetrologyAtomic Frequency StandardsClock StabilityDirect Digital SynthesisFrequency StabilityFrequency SynthesizersFrequency TuningResonanceSpectral Characteristics
New scheme of the microwave – excitation signal for the caesium atomic clock is based on method of direct digital synthesis. The theoretical calculations and experimental research showed decrease step frequency tuning by several orders and improvement the spectral characteristics of the output signal of frequency synthesizer. A range of generated output frequencies is expanded, and the possibility of detuning the frequency of the neighboring resonance of spectral line that makes it possible to adjust the C-field in quantum frequency standard is implemented. Experimental research of the metrological characteristics of the quantum frequency standard on the atoms of caesium – 133 with new design scheme of the microwave – excitation signal showed improvement in daily frequency stability on 1.2*10-14.