Thin, Soft, Skin‐Mounted Microfluidic Networks with Capillary Bursting Valves for Chrono‐Sampling of Sweat

Jungil Choi, Daeshik Kang, Seungyong Han, Sung Bong Kim, John A. Rogers

Advanced Healthcare Materials · 2017 · 305 citations · 55 references

Concepts

TL;DR

Time‑sequential capture of microliter sweat volumes from specific skin regions enables monitoring of temporal electrolyte and biomarker changes, but existing absorbent pads and electronic wearables fail to manage sweat flow for quantitative tracking. This study introduces a thin, soft, skin‑like microfluidic platform that bonds to the skin to collect and store sweat in interconnected microreservoirs. Sweat‑gland pressure drives flow through microchannels equipped with capillary bursting valves that open at distinct pressures, passively guiding sweat sequentially into the reservoirs. The device recovers 1.8 µL of sweat per 0.8 min from a 0.03 cm² skin area (~5 glands), yielding a sweat rate of 0.60 µL min⁻¹ per gland, and human studies confirm accurate temporal analysis of lactate, sodium, and potassium concentrations.

Abstract

Systems for time sequential capture of microliter volumes of sweat released from targeted regions of the skin offer the potential to enable analysis of temporal variations in electrolyte balance and biomarker concentration throughout a period of interest. Current methods that rely on absorbent pads taped to the skin do not offer the ease of use in sweat capture needed for quantitative tracking; emerging classes of electronic wearable sweat analysis systems do not directly manage sweat‐induced fluid flows for sample isolation. Here, a thin, soft, “skin‐like” microfluidic platform is introduced that bonds to the skin to allow for collection and storage of sweat in an interconnected set of microreservoirs. Pressure induced by the sweat glands drives flow through a network of microchannels that incorporates capillary bursting valves designed to open at different pressures, for the purpose of passively guiding sweat through the system in sequential fashion. A representative device recovers 1.8 µL volumes of sweat each from 0.8 min of sweating into a set of separate microreservoirs, collected from 0.03 cm 2 area of skin with approximately five glands, corresponding to a sweat rate of 0.60 µL min −1 per gland. Human studies demonstrate applications in the accurate chemical analysis of lactate, sodium, and potassium concentrations and their temporal variations.

References

55