Platelets · 2017 · 25 citations · 17 references
MedicineHematologyGood Laboratory PracticeHematology PhysiologyHemostasisPlatelet ActivationPlatelet AggregationPlatelet FunctionPlatelet PhysiologyHematocrit EffectsPlatelet Activation AssaysPlatelet Aggregometry
= 0.14). Closure time on PFA-100 testing was inversely correlated with hematocrit for both cartridges. Regression analysis revealed different 95% reference intervals in comparison with originally established intervals for both MEA and PFA-100 in low platelet or hematocrit conditions. Multiple regression analysis of ASPI and both tests on the PFA-100 for combined low platelet and hematocrit conditions revealed that only PFA-100 testing should be adjusted for both thrombocytopenia and anemia. 95% reference intervals were calculated using multiple regression analysis. However, coefficients of determination of PFA-100 were poor, and some variance remained unexplained. Thus, in this pilot study using (multiple) regression analysis, we could establish reference intervals of platelet function in anemia and thrombocytopenia conditions on PFA-100 and in thrombocytopenia conditions on MEA.
17
Marco Cattaneo, Chiara Cerletti, Paul Harrison et al. · Journal of Thrombosis and Haemostasis · 2013 · 547 citations
Medicine, Hematology, Hemostasis +11
Characterization of an In Vitro Platelet Function Analyzer, PFA-100™
Sourav K. Kundu, E Heilmann, Reynaldo Sio et al. · Clinical and Applied Thrombosis/Hemostasis · 1996 · 173 citations
Medicine, Hematology, Hemostasis +16