Multi-resolution statistical image reconstruction for mitigation of truncation effects: application to cone-beam CT of the head

Hao Dang, J. Webster Stayman, Alejandro Sisniega, Wojciech Zbijewski, Jennifer Xu, Xiaohui Wang, David H. Foos, Nafi Aygün, Vassilis E. Koliatsos, Jeffrey H. Siewerdsen

Physics in Medicine and Biology · 2016 · 23 citations · 39 references

Concepts

Abstract

voxels). Algorithm parameters (e.g. regularization strength, the ratio of the fine and coarse voxel size, and RFOV size) were investigated to guide reliable parameter selection. The findings provide a promising method for truncation artifact reduction in CBCT and may be useful for other MBIR methods and applications for which truncation is a challenge.