Interlaboratory Variability in Plasma Creatinine Measurement and the Relation with Estimated Glomerular Filtration Rate and Chronic Kidney Disease Diagnosis

Elizabeth Lee, Christine Collier, Christine A. White

Clinical Journal of the American Society of Nephrology · 2016 · 41 citations · 15 references

DOIFull text

Open access

Abstract

Significant interlaboratory variability in Cr measurement still exists. Jaffe assays yield higher Cr values than enzymatic assays, leading to lower eGFR-EPIs and more frequent CKD diagnoses. Further improvements in assay performance are required to standardize patient CKD diagnosis and to facilitate longitudinal Cr monitoring across laboratories.

References

15