A PET scatter correction using simultaneous acquisitions with low and high lower energy thresholds
58
Citations
12
References
2005
Year
Unknown Venue
Image ReconstructionEngineeringMeasurementPet-mriAdvanced ImagingPositron Emission TomographyHigh LldSimultaneous AcquisitionsCalibrationScatter CorrectionBiostatisticsComputational ImagingInstrumentationRadiation ImagingNuclear MedicineRadiologyHealth SciencesRadiation DetectionMedical ImagingSynthetic Aperture RadarRadiometrySignal ProcessingCorresponding SinogramRadarBaseline CorrectionBiomedical ImagingPet Scatter Correction
The authors propose a new method for scatter correction using two sinograms acquired simultaneously with a low and a high lower energy threshold (LLD). The high LLD is chosen such that the corresponding sinogram contains no or few scatter counts. The information in each data set are combined to produce a sinogram corrected for scatter. The method is described in detail and evaluated on various phantom measurements.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>