2013 · 11 citations · 3 references
EngineeringBioengineeringBiomedical EngineeringMicroelectromechanical SystemsNeural EngineeringUltrasoundBiomedical MemsImplantable Biomedical DevicesImplantable MicrodevicesImplantable MicrosystemsUltrasonicsWireless Power TransferBiomedical ImplantsBiomedical Wireless Power TransferCapacitive Micromachined Ultrasonic TransducersFront-end DesignImplantable Wireless Power TransferPower EfficiencyUltrasonic TransducersWireless Power Transfer For Implants
Wireless implantable devices have revolutionized the field of biomedical engineering for as long as adequate power supplies are conjured. This paper presents an ultrasonic power link front-end interfaced with a capacitive micromachined ultrasonic transducer (CMUT). The ultrasonic power link front-end consists of a rectifier, a charge pump, a clock extractor and a phase generator. The power link front-end designed in 0.18-μm CMOS process provides a 17 V DC supply for the implant microsystem, e.g. neural stimulator. It achieves an overall power efficiency of 0.3% in simulation.
3