Tremor Frequency Assessment by iPhone® Applications: Correlation with EMG Analysis
19
Citations
12
References
2016
Year
MeasurementAccelerometerWearable TechnologyDiagnosisTremor Frequency AnalysisEmg NeedleElectrophysiological EvaluationKinesiologyPatient MonitoringHealth SciencesPhysical MedicineRehabilitationMovement DisordersPhysical TherapyTremor Frequency AssessmentTremor FrequencyElectromyographyHealth MonitoringMedicine
Tremor frequency analysis is usually performed by EMG studies but accelerometers are progressively being more used. The iPhone® contains an accelerometer and many applications claim to be capable of measuring tremor frequency. We tested three applications in twenty-two patients with a diagnosis of PD, ET and Holmes' tremor. EMG needle assessment as well as accelerometry was performed at the same time. There was very strong correlation (Pearson >0.8, p < 0.001) between the three applications, the EMG needle and the accelerometry. Our data suggests the apps LiftPulse®, iSeismometer® and Studymytremor® are a reliable alternative to the EMG for tremor frequency assessment.