First application of superconducting transition-edge sensor microcalorimeters to hadronic atom X-ray spectroscopy

HEATES Collaboration, S. Okada, D. A. Bennett, C. Curceanu, W. B. Doriese, Joseph W. Fowler, Johnathon D. Gard, F. P. Gustafsson, T. Hashimoto, R. Hayano,

Progress of Theoretical and Experimental Physics · 2016 · 32 citations · 25 references

DOIFull text

Open access

Concepts

Abstract

High-resolution pionic atom X-ray spectroscopy was performed with an X-ray spectrometer based on a 240 pixel array of superconducting transition-edge sensor (TES) microcalorimeters at the <f>π</f>M1 beam line of the Paul Scherrer Institute. X-rays emitted by pionic carbon via the <f>4f→3d</f> transition and the parallel <f>4d→3p</f> transition were observed with a full width at half maximum energy resolution of 6.8 eV at 6.4 keV. The measured X-ray energies are consistent with calculated electromagnetic values which considered the strong interaction effect assessed via the Seki–Masutani potential for the <f>3p</f> energy level, and favor the electronic population of two filled <f>1s</f> electrons in the K-shell. Absolute energy calibration with an uncertainty of 0.1 eV was demonstrated under a high-rate hadron beam condition of 1.45 MHz. This is the first application of a TES spectrometer to hadronic atom X-ray spectroscopy and is an important milestone towards next-generation high-resolution kaonic atom X-ray spectroscopy.