Test of a Neutron Spectrometer in NIST Standard Fields

R. B. Schwartz, C.M. Eisenhauer

Radiation Protection Dosimetry · 1994 · 10 citations · 0 references

Concepts

Abstract

A neutron spectrometer was tested in four reference neutron fields at NIST. These fields were generated by a bare 252Cf source and by 252Cf moderated by D2O, H2O, Fe. The sensitivity and reproducibility of the spectrometer was tested by varying experimental parameters such as count rate and distance from the source. For each neutron field, measured spectra between 60 keV and 4.5 MeV are compared with Monte Carlo calculations. The spectrometer indicates peaks and valleys at locations predicted by calculations in the four different spectra. Also dose equivalent, kerma and fluence rates derived from measured spectra are compared with calculated values and agree within a few per cent when adjustments are made for energy ranges not measured by the spectrometer.