Indirect Calorimetry in Mechanically Ventilated Patients: A Prospective, Randomized, Clinical Validation of 2 Devices Against a Gold Standard
20
Citations
20
References
2016
Year
The QUARK RMR ICU Indirect Calorimeter compared better with the gold standard for values of [Formula: see text]O<sub>2</sub> and REE than did the CCM Express Indirect Calorimeter in mechanically ventilated patients who were circulatory and respiratory stable. Both indirect calorimeters had low precision.