The International Journal of Artificial Organs · 1978 · 10 citations · 0 references
MedicineBiochemistryNephrologyDialysisHemodialysisRenal Replacement TherapyDialysate CompositionRenal Blood FlowRenal ClearanceRenal Function TestingRenal Function TestsRenal Mineral HandlingContinuous MonitoringDialysis-related Drug MonitoringPhosphate RemovalUrea Kinetics
Ion-selective eletrodes have been used to monitor urea and inorganic phosphate removal during dialysis. The dialysate outlow concentrations from a single pass beside controller were used to compute instantaneous serum concentrations. The urea data are in agreement with a single pool reservoir model; the inorganic phosphate removal was not log-linear, indicating rate limiting intracompartmental transfer. The procedure can be applied to the computation of urea generation rates and thus provide information on the protein catabolism rates of the patient; in addition, the monitor provides continuous checks on the dialyzer performance.