Design-related and largely abrasion-induced tearing caused failures of this cohort of photooxidized pericardial valves. Nevertheless, this nonglutaraldehyde-preserved photofixed pericardial tissue from valves suffering design-related cuspal tears to two years postoperatively remained without significant degradation, inflammation, infection, thrombus, pannus or calcification.