Optics Letters · 2015 · 95 citations · 8 references
EngineeringChemistryPhysicsOpticsPhotonicsSensorsGas SensorsOptical InstrumentationOptical SensingOptical SensorsPressure MeasurementPressure SensorsFabry-perot ResonatorsOptical Path LengthOptical Pressure SensingOptical Pressure SensorsPressure Sensor ResolutionResonant Pressure SensorsSpacer DesignVirial Coefficients
We have built and characterized a refractometer that utilizes two Fabry-Perot cavities formed on a dimensionally stable spacer. In the typical mode of operation, one cavity is held at vacuum, and the other cavity is filled with nitrogen gas. The differential change in length between the cavities is measured as the difference in frequency between two helium-neon lasers, one locked to the resonance of each cavity. This differential change in optical length is a measure of the gas refractivity. Using the known values for the molar refractivity and virial coefficients of nitrogen, and accounting for cavity length distortions, the device can be used as a high-resolution, multi-decade pressure sensor. We define a reference value for nitrogen refractivity as n-1=(26485.28±0.3)×10(-8) at p=100.0000 kPa, T=302.9190 K, and λ(vac)=632.9908 nm. We compare pressure determinations via the refractometer and the reference value to a mercury manometer.
8
Polarizability of Helium and Gas Metrology
James W. Schmidt, R. M. Gavioso, Eric F. May et al. · Physical Review Letters · 2007 · 147 citations
Natural Sciences, Chemistry, Physics +12
Absolute refractometry of dry gas to ±3 parts in 10^9
Patrick Egan, Jack A. Stone · Applied Optics · 2011 · 73 citations
Leslie Pendrill · Metrologia · 2004 · 55 citations
Natural Sciences, Chemistry, Physics +17