First Results From a High-Resolution Small Animal SiPM PET Insert for PET/MR Imaging at 7T

Andrew L. Goertzen, Greg Stortz, Jonathan D. Thiessen, D. Bishop, Muhammad Salman Khan, Piotr Kozłowski, F. Retière, Graham Schellenberg, Ehsan Shams, Vesna Sossi,

IEEE Transactions on Nuclear Science · 2016 · 52 citations · 29 references

Concepts

Abstract

We present the initial results from a small animal PET insert designed to be operated inside a 7T MRI. The insert fits within the 114 mm inner diameter of the Bruker BGA-12S gradient coil while accommodating the Bruker 35 mm volume RF coil (outer diameter 60 mm), both used in the Bruker 70/20 MRI systems. The PET insert is a ring comprising 16 detectors. Each detector has a dual-layer offset (DLO) lutetium-yttrium oxyorthsilicate (LYSO) scintillator array read out by two SensL SPMArray4B SiPM arrays. The DLO scintillator has bottom (top) layers of: <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$22\times 10$ </tex-math></inline-formula> ( <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$21 \times 9$ </tex-math></inline-formula> ) crystals of size <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$1.2 \times 1.2 \times 6$ </tex-math></inline-formula> (4) mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">3</sup> for a total of 409 crystals per block, providing an axial extent of 28.17 mm. The detector outputs are multiplexed to four signals using a custom readout board and digitized using the OpenPET data acquisition platform. The detector flood images successfully resolve over 99% of the crystals, with average energy resolution of 12.5 ± 2.0% at 511 keV. Testing of the PET system inside the MRI showed that the PET insert had no effect on MRI image homogeneity and only a small effect on echo planar images (EPI) signal to noise ratio (SNR) (-9%), with neither PET nor MRI images showing obvious artefacts. These acquisitions used the OpenPET operating in “oscilloscope mode” with USB2.0 interface, allowing a maximum total singles event rate of 280 kcps, strongly limiting the count rate capabilities of the system. The PET radial spatial resolution (as measured with a <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">22</sup> Na point source and FBP-3DRP reconstruction) is 1.17 mm at the centre, degrading to 1.86 mm at a 15 mm radial offset. Simultaneous phantom and mouse PET/MR imaging produced good quality images that were free of any obvious artefacts.

References

29