No evidence for endotoxin transfer across high flux polysulfone membranes.

J. Bommer, Becker Kp, R. Urbaschek, Eberhard Ritz, B. Urbaschek

PubMed · 1987 · 59 citations · 0 references

Concepts

Abstract

Recently, there has been some concern that high-flux membranes may expose dialysis patients to the risk of endotoxin transfer secondary to backfiltration within the dialyzer. To evaluate the safety of high-flux polysulfone dialyzers, we examined in an in vitro recirculation system whether lipopolysaccharides (LPS) and lipid A respectively penetrate from the dialysate to the blood compartment and vice versa using a F-60 hemofilter (Fresenius AG). For the detection of endotoxin, a sensitive, kinetic limulus amebocyte lysate (LAL) microtiter test was used. It can be concluded that LPS and lipid A do not pass from either side through the filter system used when saline was recirculated for more than 10 h on both sides of the membrane.