Gamma Ray Measurements at OMEGA with the Newest Gas Cherenkov Detector “GCD-3”

13

Citations

8

References

2016

Year

Abstract

Initial results from the newest Gas Cherenkov Detector (GCD-3) are reported demonstrating improved performance over previous GCD iterations. Increased shielding and lengthening of the Cherenkov photon optical path have resulted in a diminished precursor signal with increased temporal separation between the precursor and the primary DT Cherenkov signal. Design changes resulted in a measured GCD-3 sensitivity comparable to GCD-1 at identical 100 psia CO<sub>2</sub> operation. All metal gasket seals and pressure vessel certification to 400 psia operation allow for a GCD-3 lower Cherenkov threshold of 1.8 MeV using the fluorinated gas C<sub>2</sub>F<sub>6</sub> as compared to the 6.3 MeV lower limit of GCD-1 and GCD-2. Calibration data will be used to benchmark GEANT4 and ACCEPT detector models. Lastly, the GCD-3 acts as a prototype for the Super GCD being fielded at the National Ignition Facility (NIF) as part of the National Diagnostics Plan and will be installed at NIF in early 2016.