Japanese Journal of Applied Physics · 2016 · 20 citations · 26 references
EngineeringBioengineeringBiomedical EngineeringElectrophysiologyMicrofabricationNeural EngineeringElectroencephalographyImplantable Biomedical DevicesImplantable MicrodevicesImplantable MicrosystemsMicrofabrication MaterialsDry ElectrodesElectrode - Skin InterfaceElectrode-skin InterfaceMicroneedle FabricationMicroneedle Geometry DesignMicroneedlesPolymeric MicroneedlesWet Electrodes
Abstract In this paper, we report on the optimization of the shape of dry microneedle electrodes for electroencephalography (EEG) on hairy locations and compare the electrodes we developed with conventional wet electrodes. We propose the use of SU-8-based candle-shaped microneedle electrodes (CMEs), which have pillars of 1.0 mm height and 0.4 mm diameter with a gap of 0.43 mm between pillars. Microneedles are formed on the top of the pillars. The shape was determined by how well the pillars can avoid hairs and support the microneedles to penetrate through the stratum corneum. The skin–electrode contact impedances of the fabricated CMEs were found to be higher and less stable than those of conventional wet electrodes. However, the CMEs successfully acquired signals with qualities as good as those of conventional wet electrodes. Given the usability of the CMEs, which do not require skin preparation or gel, they are promising alternatives to conventional wet electrodes.
26
<i>Electric Fields of the Brain: The Neurophysics of EEG</i>
Paul L. Nunez, E. Harth · Physics Today · 1982 · 2.3K citations
Scalp electrode impedance, infection risk, and EEG data quality
Thomas C. Ferrée, Phan Luu, Gerald S. Russell et al. · Clinical Neurophysiology · 2001 · 1K citations