Iranian Journal of radiation research/Iranian journal of radiation research · 2015 · 12 citations · 10 references
Radiation MedicineMedical ImagingRadiation TherapyCommercial TreatmentSquare Fi EldBrachytherapy3-D ConformalDosimetryRadiaon TherapyRadiation Therapy PlanningProton TherapyInstrumentationSynchrotron RadiationMedicineRadiation OncologyTreatment VerificationRadiologyHealth Sciences
Background: The present work aims to study the out-of-field photon calculaon accuracy of a commercial treatment planning system (TPS), Oncentra, for three-dimensional conformal radiotherapy (3D-CRT) and intensity modulated radiaon therapy (IMRT) treatment techniques from Elekta Synergy medical linear accelerator. Materials and Methods: Accuracy of individual out-of-field dose components was studied by defining a square fi eld of 10 x 10 cm 2
10
A Monte Carlo model for out‐of‐field dose calculation from high‐energy photon therapy
Stephen F. Kry, U Titt, D Followill et al. · Medical Physics · 2007 · 93 citations