Insight into the quantitative metrics of chemical exchange saturation transfer (CEST) imaging

Hye‐Young Heo, Dong‐Hoon Lee, Yi Zhang, Xüna Zhao, Shanshan Jiang, Min Chen, Jinyuan Zhou

Magnetic Resonance in Medicine · 2016 · 95 citations · 69 references

Concepts

Abstract

Purpose To evaluate the reliability of four CEST imaging metrics for brain tumors, at varied saturation power levels and magnetic field strengths (3−9.4 Tesla (T)). Methods A five‐pool proton exchange model (free water, semisolid, amide, amine, and NOE‐related protons) was used for the simulations. For the in vivo study, eight glioma‐bearing rats were scanned at 4.7 T. The CEST ratio (CESTR), CESTR normalized with the reference value (CESTR nr ), inverse Z‐spectrum‐based (MTR Rex ), and apparent exchange‐related relaxation (AREX) were compared. Results The simulated CEST signal intensities using MTR Rex and AREX were substantially increased at relatively high radiofrequency (RF) saturation powers at 3 T and 4.7 T, whereas CESTR and CESTR nr metrics remained relatively stable. There were tremendously high MTR Rex and AREX signals around the water frequency at all field strengths because of the small denominators. In the rat tumor study at 4.7 T, both CESTR and CESTR nr showed clear contrasts in the tumor with respect to the normal tissue across all saturation power levels (0.5−3 μT), whereas the AREX showed negligible to negative insignificant contrasts. Conclusions CEST metrics must be carefully selected based on the different experimental settings. CESTR and CESTR nr are more reliable at 3 T (a clinical field strength) and 4.7 T. Magn Reson Med 77:1853–1865, 2017. © 2016 International Society for Magnetic Resonance in Medicine