Factors That Affect Corneal Flap Thickness With the Hansatome Microkeratome

Raymond P. Gailitis, Mara Lagzdins

Journal of Refractive Surgery · 2002 · 57 citations · 16 references

Abstract

The Hansatome tended to cut thinner flaps than anticipated based on the plate used. Flaps cut on the first eye were thicker than the second eye using the same blade. Thicker corneas tended to lead to thicker flaps. There was no correlation between flap thickness and microkeratome ring size, temperature, barometric pressure, patient age, or average keratometric power.