Operation of 1020-MHz NMR Superconducting Magnet

G. Nishijima, Shinji Matsumoto, Kenjiro Hashi, Shinobu Ohki, Atsushi Goto, Takashi Noguchi, Shuji Sakai, Tadashi Shimizu

IEEE Transactions on Applied Superconductivity · 2016 · 11 citations · 8 references

Concepts

Abstract

We had successfully operated the 1020-MHz (24.0 T) nuclear magnetic resonance (NMR) spectrometer for 8 months, including the cooling down, shimming, and the 6-month operation for NMR measurements. The NMR superconducting magnet, which had consisted of NbTi and Nb <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">3</sub> Sn coils, was upgraded by replacing the innermost Nb <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">3</sub> Sn coil with a (Bi,Pb) <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> Sr <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> Ca <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> Cu <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">3</sub> O <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">10</sub> (Bi-2223) one. The Bi-2223 innermost coil was connected in series with Nb <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">3</sub> Sn and NbTi coils. The magnet was cooled with pressurized superfluid helium (He II). The steady-state operating temperature was 1.72 K. The magnet was power-supply driven, i.e., the transport current was always supplied by a power supply. A small normal transition and a power outage occurred during the 8-month operation; however, the magnet quench was avoided.

References

8