IEEE Photonics Technology Letters · 2016 · 28 citations · 18 references
EngineeringBioengineeringBiomedical EngineeringPhotonicsSensorsSpectroscopyAbsorption SpectroscopyInfrared SpectroscopyLaser SpectroscopyOptical BiosensingOptical SensorsOptical SpectroscopyInfrared Absorption SpectroscopyInfrared Gas SensorsInstrument Detection LimitMethaneMethane AbsorptionMethane DetectionMethod Detection LimitTime Resolution
A hollow waveguide (HWG)-enhanced tunable laser absorption spectroscopy sensor was demonstrated for the detection of exhaled methane in the mid-infrared range. A 3392-nm interband cascade laser was used to simultaneously target two neighboring strong absorption peaks and wavelength modulation spectroscopy (WMS) with the second harmonic signal normalized by the first-harmonic signal (WMS- <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$2f/1f$ </tex-math></inline-formula> ) was employed for data processing. With a time resolution of 0.1 s and a detection limit of 48 ppb, the evolution of exhaled methane concentration in a single breath was observed. In addition, the HWG-based optical path is mirror-less, flexible, and light, which makes the proposed sensor suited for clinical instrumentation, point-of-care breath monitoring, and disease diagnostics.
18