Concepedia

Publication | Closed Access

High-Salt-Tolerance Matrix for Facile Detection of Glucose in Rat Brain Microdialysates by MALDI Mass Spectrometry

106

Citations

21

References

2011

Year

Abstract

Due to its strong ultraviolet absorption, high salt tolerance, and little interference in the low molecular weight region, N-(1-naphthyl) ethylenediamine dihydrochloride (NEDC) has been applied as a matrix to measure the level of glucose in rat brain microdialysates by matrix-assisted laser desorption/ionization mass spectrometry (MALDI MS) in combination with in vivo microdialysis. By monitoring the ion signals of (glucose + Cl)(-) in the mass spectra, we achieved a low detection limit of ~10 μM for glucose in 126 mM NaCl, which is a typical component in artificial cerebrospinal fluid, without prior sample purification. It is concluded that NEDC-assisted laser desorption/ionization (LDI) MS is a fast and general method for sensitive detection of small molecules (such as glucose and amino acids) in high ionic strength solutions.

References

YearCitations

Page 1