Analytical Chemistry · 2011 · 107 citations · 21 references
Life SciencesBiochemistryNeuroscienceAnalytical ChemistryMass Spectrometric AnalysisNeurochemistryBioanalytical Method ValidationMass AnalyzersMass SpectrometryAmino AcidsGlucose SensingInstrument Detection LimitIonic SolutionsIonic StrengthMatrix ApplicationMatrix-assisted Laser Desorption/ionizationSalt Tolerance
Due to its strong ultraviolet absorption, high salt tolerance, and little interference in the low molecular weight region, N-(1-naphthyl) ethylenediamine dihydrochloride (NEDC) has been applied as a matrix to measure the level of glucose in rat brain microdialysates by matrix-assisted laser desorption/ionization mass spectrometry (MALDI MS) in combination with in vivo microdialysis. By monitoring the ion signals of (glucose + Cl)(-) in the mass spectra, we achieved a low detection limit of ~10 μM for glucose in 126 mM NaCl, which is a typical component in artificial cerebrospinal fluid, without prior sample purification. It is concluded that NEDC-assisted laser desorption/ionization (LDI) MS is a fast and general method for sensitive detection of small molecules (such as glucose and amino acids) in high ionic strength solutions.
21
Colorimetric Detection of Glucose in Rat Brain Using Gold Nanoparticles
Ying Jiang, Hong Zhao, Yuqing Lin et al. · Angewandte Chemie International Edition · 2010 · 274 citations
Bianzhi Xing, Hui Chen, Min Zhang et al. · Stroke · 2008 · 247 citations
Medicine, Neurobiology, Neuroscience +15