IEEE Sensors Journal · 2014 · 21 citations · 32 references
EngineeringMechanical EngineeringHeat TransferSensorsSurface EngineeringSensor CalibrationTemperature MeasurementTemperature MetrologyTemperature SensorsThermal ManagementThermal SensorsHeat Flux SensingTemperature CalibrationTemperature DependenceTemperature SensingTemperature Sensor SensitivityThermal CalibrationThermal Sensor CalibrationTransient Thermal Response
Heat flux sensors fabricated using Direct Write Thermal Spray are thin, surface-based devices that can operate at high temperatures. In this paper, Direct Write heat flux sensors are calibrated over a temperature range of 25 °C-860 °C. A substitution-based quartz lamp configuration is used to measure the steady-state sensitivity and transient response of Direct Write sensors at ambient temperatures, with repeatability confirmed over a 10-month period and after thermal aging. A matched heat flow approach is used to characterize the sensors at operating temperatures up to 860 °C. The sensitivity is found to increase by a factor of two from 25 °C to 650 °C, after which it plateaus up to the maximum tested temperature of 860 °C. A cubic polynomial calibration function captures the temperature dependence of the sensitivity and provides a good agreement for the measured data.
32
P. Fauchais, Ghislain Montavon, R.S. Lima et al. · Journal of Physics D Applied Physics · 2011 · 218 citations