Journal of research of the National Bureau of Standards · 1945 · 23 citations · 6 references
Natural SciencesMaterials EngineeringInstrumentationMetrologyOpticsInterferometryOptical InstrumentsOptical MetrologyRefractive IndexFringe AnalysisInterference Fringe AnalysisInterferometric Expansion MeasurementInterferometric MeasurementInterferometric Strain MeasurementOptical Interferometry
A photographic instrument is described that was designed for recording the changes produced in the order of interfer ence fringes over long periods and for recording large changes in the order of interference. When this instrument is used with interferometric systems for studying the changes caused in transparent solids by heating, it can be made to yield a continuous record of the simultaneous changes in temperature, time, index, strain, and density. For example, to record those changes that are caused by the annealing of an optical glass at an annealing temperature near the lower part of its annealing range requires several weeks of continuous r ecording. The total amount of 35-millimeter film needed by the instrument for such a record (covering, say 3 months) does not necessarily exceed 10 feet. To record the expansion of a material that is being heated from room t emperature to 500 0 C, at a rate of 3 degrees centigrade per minute, requires approximately 1 foot of 35-millimeter film. Thus, even very long tests r equire such a small amount of film that t h e operator is able to process it in any small dark room. The fringes a re photographed at t heir natural size, and the fringe shifts can be determined from the photographic record without the use of enlarging or projection equipment.
6