Comparison of Heat Transfer Measurement Devices in Arc Jet Flows with Shear

Imelda Terrazas-Salinas, Jose A. Carballo, David Driver, John Balboni

10th AIAA/ASME Joint Thermophysics and Heat Transfer Conference · 2010 · 29 citations · 12 references

Concepts

Abstract

*† ‡ This paper presents results of an arc jet test campaign and analyses to identify the sources of measurement errors. The primary objective of the tests was to evaluate different instruments available for measuring convective heat flux in an arc-heated gas flow to wedge test bodies in a shear environment and at the stagnation point of blunt cylinders. The campaign evaluated four types of heat flux sensors: 1) slug calorimeters, 2) Gardon gages, 3) null-point calorimeters, and 4) thin-skin calorimeters. Results from each of the devices agreed reasonably well in stagnation flow, but not so well in shear flow. We performed numerical modeling of the responses of the various devices in an effort to better understand the differences and to quantify the errors. Gardon gages with dimples in them were found to under-report the heat transfer by 6%. Gaps on slug calorimeters were found to admit small amounts of side-wall heating that contributed on the order of 1% to the overall heating on the slug. Null-point calorimeters agreed with thin-skin calorimeters in stagnation, but not in shear environments. We explored catalytic heating effects: The results seem to indicate that the various heat transfer devices may not be as catalytic as we had previously thought.

References

12