Double-Focusing Thermal Triple-Axis Spectrometer at the NCNR
118
Citations
13
References
2012
Year
EngineeringNuclear PhysicsNuclear DataOptical PropertiesInfrared OpticNuclear MaterialsInstrumentationRadiation ImagingNuclear ReactorsDiffraction DetectorRadiologyRadiation DetectionPhysicsHorizontal EnergyAccelerator Mass SpectrometryNuclear SecurityNeutron SourceExperimental Nuclear PhysicsNatural SciencesSpectroscopyInstrument ScienceApplied PhysicsDetector PhysicNeutron ResearchNeutron ScatteringSpectroscopic Method
The new thermal triple-axis spectrometer at the NIST Center for Neutron Research (NCNR) is located at the BT-7 beam port. The 165 mm diameter reactor beam is equipped with a selection of Söller collimators, beam-limiters, and a pyrolytic graphite (PG) filter to tailor the beam for the dual 20×20 cm(2) double-focusing monochromator system that provides monochromatic fluxes exceeding 10(8) n/cm(2)/s onto the sample. The two monochromators installed are PG(002) and Cu(220), which provide incident energies from 5 meV to above 500 meV. The computer controlled analyzer system offers six standard modes of operation, including a diffraction detector, a position-sensitive detector (PSD) in diffraction mode, horizontal energy focusing analyzer with detector, a Q-E mode employing a flat analyzer and PSD, a constant-E mode with the analyzer crystal system and PSD, and a conventional mode with a selection of Söller collimators and detector. Additional configurations for specific measurement needs are also available. This paper discusses the capabilities and performance for this new state-of-the-art neutron spectrometer.