Study of High-Resolution Mass Spectrometry Technology as a Replacement for Tandem Mass Spectrometry in the Field of Quantitative Pesticide Residue Analysis

Anton Kaufmann, Veronika Dvorak, Claudio Crüzer, Patrick Butcher, Kathryn Maden, Stephan Walker, Mirjam Widmer, Andreas Schürmann

Journal of AOAC International · 2012 · 40 citations · 0 references

DOIFull text

Open access

Concepts

Abstract

A validated LC/MS/MS-based multiresidue pesticide method was converted to an LC high-resolution MS single-stage Orbitrap platform. No changes regarding the cleanup and LC were made. Optimization of high-resolution MS-specific parameters and interface settings was kept at a minimum. The aim was to explore the capability of current Orbitrap technology to substitute for LC/MS/MS technology. The test included the quantitative performance (sensitivity, selectivity, linearity, accuracy, and precision) of some 240 analytes in three different matrixes. The LC/MS/MS instrumentation was operated at the edge of its technical limitations. A further extension of the number of analytes for LC/MS/MS would require the use of even narrower dwell times, significantly reducing sensitivity and reproducibility of measurement. No such limitations exist for the high-resolution technology. Similar performance was observed for both technologies. A current drawback of the high-resolution technology is the speed of data processing, which took significantly longer than for LC/MS/MS data due to the limited capabilities of the software.