Publication | Closed Access
Transcutaneous oxygen monitoring
14
Citations
0
References
1985
Year
Heart FailurePulmonary CirculationCardiovascular DiseaseTranscutaneous Oxygen MonitoringSkin SurfacePatient MonitoringBlood Flow MeasurementPerioperative MonitoringTissue OxygenationBiomedical EngineeringCutaneous Blood FlowMedicineCardiologyEmergency MedicineAnesthesiologyBlood Flow
The technology to measure noninvasively the partial pressure of oxygen at the skin surface (transcutaneous PO2, or PtcO2) is now commercially available. In patients with normal cardiac output and cutaneous blood flow, the PtcO2 accurately monitors changes in the arterial partial pressure of oxygen (PaO2). However, if the cardiac output is reduced, the PtcO2 diverges from PacO2 and reflects tissue oxygen delivery instead. Thus, the physiologic interpretation of the PtcO2 varies according to the hemodynamic status of the patient. This limits the utility of transcutaneous oxygen monitoring in critically ill patients, but such monitoring can be useful in the noninvasive detection of adverse changes in arterial oxygenation or tissue perfusion.