Comparison of four automated serum vitamin B12 assays
46
Citations
2
References
2015
Year
Total B12 assays correlate very well with each other. However, results of DxI 800 Unicel were lower compared to the other three autoanalyzers. All total vitamin B12 methods show similar relationships with HCY and MMA. Standardization of serum vitamin B12 assays is still not completed and further standardization studies are needed. Laboratory professionals and clinicians should be aware of this disagreement between assay methods and they should use these tests as ancillary tests.