SQUID Microscope With Hollow-Structured Cryostat for Magnetic Field Imaging of Room Temperature Samples

Jun Kawai, Hirokuni Oda, J. Fujihira, Masakazu Miyamoto, Isoji Miyagi, Masahiko Sato

IEEE Transactions on Applied Superconductivity · 2016 · 25 citations · 12 references

Concepts

Abstract

We have developed a high-spatial-resolution superconducting quantum interference device (SQUID) microscope for magnetic field imaging of rock samples at room temperature, using a hollow-structured cryostat. A directly coupled low-temperature SQUID with a 200 μm × 200 μm pickup loop, which is mounted on a sapphire conical rod, is separated from room temperature by a thin sapphire window. Precise and repeatable adjustment of the vacuum gap between the SQUID and the sapphire window is performed by rotating a micrometer spindle connected to the sapphire rod through the hollow portion of the cryostat. A spacing of 230 μm between the SQUID and a sample has been achieved. We demonstrated the imaging of the magnetic field on a zircon containing magnetic grains.

References

12