Publication | Open Access
Variability and bias assessment in breast ADC measurement across multiple systems
33
Citations
22
References
2016
Year
This breast phantom can be used to measure scanner-coil-sequence bias and variability for DWI. When establishing a multisystem study, this breast phantom may be used to minimize protocol differences (e.g., due to available sequences or shimming technique), to correct for bias that cannot be minimized, and to weigh results from each system depending on respective variability. J. Magn. Reson. Imaging 2016. J. MAGN. RESON. IMAGING 2016;44:846-855.