Evaluation of adhesion promoters for Parylene C on gold metallization

V. Radun, Rene P. von Metzen, Thomas Stieglitz, Volker Bucher, Alfred Stett

Current Directions in Biomedical Engineering · 2015 · 15 citations · 5 references

DOIFull text

Open access

Concepts

Abstract

Abstract Delamination of thin film polymeric coatings from metallization layers is a common cause of failure in biomedical implants. To address the problem, different adhesion promotion techniques can be applied which include surface pre-treatment with oxygen and argon plasma and the use of different adhesion promoters. In this paper the applicability of titanium (Ti), silicon oxide (SiO x ), diamond-like carbon (DLC), tetramethylsilane (TMS) and aluminium oxide (AlO x ) as adhesion promoters is evaluated. A cross cut, peel and scratch test are used to qualify and quantify the adhesion before and after storage in phosphate buffered saline (PBS) for 48 hours at a temperature of 37 °C. Promising results could be achieved by a combination of Ti and DLC as well as by AlO x .

References

5