Accuracy of heart rate variability estimation by photoplethysmography using an smartphone: Processing optimization and fiducial point selection

V. Ferrer-Mileo, Federico Guede-Fernández, Mireya Fernández-Chimeno, J. Ramos-Castro, Miguel A. García-González

2015 · 15 citations · 8 references

DOIFull text

Open access

Concepts

Abstract

This work compares several fiducial points to detect the arrival of a new pulse in a photoplethysmographic signal using the built-in camera of smartphones or a photoplethysmograph. Also, an optimization process for the signal preprocessing stage has been done. Finally we characterize the error produced when we use the best cutoff frequencies and fiducial point for smartphones and photopletysmograph and compare if the error of smartphones can be reasonably be explained by variations in pulse transit time. The results have revealed that the peak of the first derivative and the minimum of the second derivative of the pulse wave have the lowest error. Moreover, for these points, high pass filtering the signal between 0.1 to 0.8 Hz and low pass around 2.7 Hz or 3.5 Hz are the best cutoff frequencies. Finally, the error in smartphones is slightly higher than in a photoplethysmograph.