Concepedia

Publication | Closed Access

Comprehensive Analysis of a Recalled Modular Total Hip System and Recommendations for Management

54

Citations

41

References

2016

Year

Abstract

Early failures of modular total hip arthroplasty occur due to fretting and corrosion at the neck-stem junction, resulting in ALTR. Surveillance utilizing metal ion levels and MRI may be indicated for all patients regardless of symptoms, as the early survivorship is poor and the ultimate failure rate may be catastrophically high.