Publication | Open Access
Radiography and histology of a threaded acetabular implant. One case studied at two years
39
Citations
0
References
1988
Year
Tissue EngineeringBone ImagingHip ArthroplastyMedicineDental BiomechanicsThreaded Acetabular ImplantOrthognathic SurgerySurgeryOsteoporosisAnatomyAcetabular InterfaceJoint ReplacementImplantologyOrthopaedic SurgeryThreaded Titanium RingAcetabular ComponentsRadiology
Threaded acetabular components are widely used in cementless total hip replacement, despite a poor understanding of the nature of the bone-implant interface. We have examined one case in which the threaded titanium ring appeared to be well incorporated with no discernible radiolucency. Microradiography and histology surprisingly showed that the threads were entirely encapsulated in fibrous tissue. This raises doubt about the relevance of plain radiography to the analysis of the acetabular interface.