RADIATION DOSIMETRY PROPERTIES OF SMARTPHONE CMOS SENSORS

Olivier Van Hoey, Alexia Salavrakos, António Marques, Alexandre Nagao, Ruben Willems, F. Vanhavere, V. Cauwels, Luana de Freitas Nascimento

Radiation Protection Dosimetry · 2015 · 24 citations · 2 references

Concepts

Abstract

During the past years, several smartphone applications have been developed for radiation detection. These applications measure radiation using the smartphone camera complementary metal-oxide-semiconductor sensor. They are potentially useful for data collection and personal dose assessment in case of a radiological incident. However, it is important to assess these applications. Six applications were tested by means of irradiations with calibrated X-ray and gamma sources. It was shown that the measurement stabilises only after at least 10-25 min. All applications exhibited a flat dose rate response in the studied ambient dose equivalent range from 2 to 1000 μSv h(-1). Most applications significantly over- or underestimate the dose rate or are not calibrated in terms of dose rate. A considerable energy dependence was observed below 100 keV but not for the higher energy range more relevant for incident scenarios. Photon impact angle variation gave a measured signal variation of only about 10 %.