Fusion Science & Technology · 2007 · 14 citations · 2 references
Natural SciencesRadiologyDiagnostic ImagingOpticsRadiation MeasurementRadiographic ImagingX-ray Data AnalysisX-ray DetectionX-ray OpticsOptical DepthOptical Penetration DepthSpatial Power SpectrumSpatial Resolution MeasurementX-ray Imaging PhysicsX-ray Imaging SystemsX-ray Optics Testing
We built a precision radiography system that measures shells for all current ablator materials to an accuracy of 1:104 in optical depth fluctuation and a spatial resolution of 120 μm. The data obtained by the precision radiography system for undoped shells was compared with the data taken using the well-known surface measurement technique Spheremapper. Since both techniques yielded the same power spectrum for the same shell, the results of the precision radiography system were verified. When this technique is compared to the Be:Cu NIF shell, there is no significant internal layer fluctuation. To account for the growing measurement demand, a new x-ray system to accommodate measurements in 1 working day was designed.
2
Complete Surface Mapping of ICF Shells
R. B. Stephens, D. K. Olson, H. Huang et al. · Fusion Science & Technology · 2004 · 32 citations · Full text