SAE technical papers on CD-ROM/SAE technical paper series · 2007 · 22 citations · 9 references
EngineeringElectrical EngineeringInstrumentationPerformance MeasurementAir Pollution MeasurementCarbon Emissions MeasurementEmissions TestingEmission MeasurementEmission Performance TestingOn-board Emission Measurement
<div class="htmlview paragraph">A new on-board portable emission measurement system (PEMS) for gaseous emissions has been designed and developed to meet CFR Part 1065 requirements. The new system consists of a heated flame ionization detector (HFID) for the measurement of total hydrocarbon, a heated chemiluminescence detector (HCLD) for the measurement of NOx, and a heated non-dispersive infra-red detector (HNDIR) for the measurement of CO and CO<sub>2</sub>. The oxygen interference and relative sensitivity of several hydrocarbon components have been optimized for the HFID. The CO<sub>2</sub> and H<sub>2</sub>O quenching effect on the HCLD have been compensated using measured CO<sub>2</sub> and H<sub>2</sub>O concentration. The spectral overlap and molecular interaction of H<sub>2</sub>O on the HNDIR measurement has also been compensated using an independent H<sub>2</sub>O concentration measurement. The basic performance of the new on-board emission measurement system has been verified accordingly with CFR part 1065 and all of the performances have met with CFR part 1065 requirement. The new on-board emission measurement system has been also compared with conventional laboratory instrumentation. The results showed good correlation with that of a laboratory system.</div>
9
Timothy J. Truex, John Collins, Jeff J. Jetter et al. · SAE technical papers on CD-ROM/SAE technical paper series · 2000 · 30 citations
Engineering, Automotive Engineering, Environmental Engineering +17