Studies in Derivative Chronopotentiometry: I . Instrumentation and Diffusion‐Controlled Systems

Peter E. Sturrock, J. L. Hughey, B. Vaudreuil, G.E. O'Brien, Robert H. Gibson

Journal of The Electrochemical Society · 1975 · 17 citations · 0 references

Concepts

Abstract

Theoretical correction and instrumental compensation of charging current are discussed for both conventional and derivative chronopotentiometry. Experimental results indicate that either correction or compensation is effective in reducing the error due to charging the electrical double layer, but that best results are obtained by application of both techniques. In diffusion‐controlled systems, derivative chronopotentiometry shows significant advantages over conventional chronopotentiometry.